| Capacity | 200 liter |
| Type | Open Care |
| Material | Mild Steel |
| Brand | ACCULOC |
| Usage/Application | Hospital |
| Color | Gray |
| Country of Origin | Made in India |
| Service Location | All Over India |
OPERATING MODES SKIN MODE: In Skin mode heater power is controlled with reference to Skin probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. AIR MODE: In Air mode heater power is controlled with reference to Air probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. MANUAL: MODE In manual mode user can control heater manually. In this mode user can adjust heating speed (0 to 8 speed) and set timer upto 59min. With probes installed when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. Special attention is needed while operating without
ALERTS Power fail: Power fail indicator turns ON with sound alert (cannot be muted). Probe fail: Probe fail indicator turns ON with sound alert (can be muted). Displays “-.-.-“on temp display. Heater overload: Heater overload indicator turns ON. Low temperature: Low alarm indicator turns ON with sound alert (can be muted). Displays blinks. High temperature: High alarm indicator turns ON with sound alert (can be muted). Displays blinks. Manual mode Time out: Short sound alert and timer flashing on LCD.
Additional Information:
- Production Capacity: As per Client Requirement
- Delivery Time: 2 DAYS
- Packaging Details: IN BOX,WHEEL, BROCHURE,CABLE