Mumbai, Maharashtra
GST No. 27CWGPS6694J1ZL
Call 08048953297 91% Response Rate
Minimum Order Quantity: 10 Piece
Product Brochure
Temperature Sensor | RTD - PT100 |
Relative Humidity | NA |
Baby Control | NA |
Brand | APS Embedded Systems |
Mobile Castors | NA |
Technology | PID |
Input Voltage | 220V AC |
Material | PCB |
Heater Power | 650W - 1000W |
Overall Dimensions | NA |
Heater Rotation | NA |
Model No. | BW Servo Controller V6.1 |
Wavelength | NA |
Size | NA |
Load Capacity | NA |
Height | 190mm |
Dimensions | NA |
Feature | PID control, all types of fail safe alarms |
Shape | NA |
Bed Capacity | NA |
Effective Area | NA |
Irradiance | NA |
Control output Type | Relay/SSR |
Resolution | 0.1 deg C |
Color | NA |
Accuracy | 0.2 deg C |
Fuse | 5A |
Model Name/Number | Skin Air type dual control. |
Operating Voltage | 220V |
Temperature Resolution | 0.1Deg |
Temperature | 25-38 |
Suitable For | All types of Infant Warmers |
Air Temperature | 25-38 |
Heater Wattage | 650W |
Warranty | 2 years |
Function | Accurate |
Finish | NA |
Country of Origin | Made in India |
Input Type | RTD |
Control Type | PID |
Humidity | NA |
Max Capacity | NA |
Product Code | BW5V4 |
Phase | Single |
Heating element | Ceramic heater |
Temperature Accuracy | 0.2 Deg |
Monitor Tray | NA |
Type | Controller board with RTD PT 100 skin and air temperature sensors |
Control Accuracy | High PID control |
Wattage | 650W |
Modes | Skin, Air and manual |
Display | large seven segment for actual Skin and Air, LCD for all alerts, Apgar and heater power |
Footprint | NA |
Heater Output | 650W |
Wheel Diameter | NA |
Drawer Size | NA |
Usage/Application | Hospital |
Storage Temperature | NA |
Footprint Size | NA |
Frequency | 50Hz |
Capacity | 1000W |
Mattress Size | NA |
Depth | NA |
Heater Capacity | 650W |
Width | 116mm |
Operating Temperature | 25-38 |
Voltage | 220V |
OPERATING MODES -
SKIN MODE: In Skin mode heater power is controlled with reference to Skin probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off.
AIR MODE: In Air mode heater power is controlled with reference to Air probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off.
MANUAL: MODE In manual mode user can control heater manually. In this mode user can adjust heating speed (0 to 8 speed) and set timer upto 59min. With probes installed when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. Special attention is needed while operating without
ALERTS
Power fail: Power fail indicator turns ON with sound alert (cannot be muted).
Probe fail: Probe fail indicator turns ON with sound alert (can be muted). Displays “-.-.-“on temp display.
Heater overload: Heater overload indicator turns ON.
Low temperature: Low alarm indicator turns ON with sound alert (can be muted). Displays blinks.
High temperature: High alarm indicator turns ON with sound alert (can be muted). Displays blinks.
Manual mode Time out: Short sound alert and timer flashing on LCD.
View Complete details
Minimum Order Quantity: 100 Piece
Product Brochure
Length Mm | na |
Light Source | na |
Number Of Wheels | 4 |
Temp Range Deg Celsius | 38 |
Model Name/Number | BW - 5V1 |
Brand | APS Embedded Systems |
Material | Metal |
Usage/Application | Nicu |
Color | NA |
Country of Origin | Made in India |
Voltage | 220V AC |
Height | na |
Width | na |
Temperature | Skin and Air Servo Controlled |
Type | Dual Servo Controlled skin and air |
Assay Time Min | na |
Power V | 220VAC |
OPERATING MODES SKIN MODE: In Skin mode heater power is controlled with reference to Skin probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. AIR MODE: In Air mode heater power is controlled with reference to Air probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. MANUAL: MODE In manual mode user can control heater manually. In this mode user can adjust heating speed (0 to 8 speed) and set timer upto 59min. With probes installed when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. Special attention is needed while operating without
ALERTS Power fail: Power fail indicator turns ON with sound alert (cannot be muted). Probe fail: Probe fail indicator turns ON with sound alert (can be muted). Displays “-.-.-“on temp display. Heater overload: Heater overload indicator turns ON. Low temperature: Low alarm indicator turns ON with sound alert (can be muted). Displays blinks. High temperature: High alarm indicator turns ON with sound alert (can be muted). Displays blinks. Manual mode Time out: Short sound alert and timer flashing on LCD.
Additional Information:
View Complete details